R&S®SML / R&S®SMV03 List of Error Messages
1090.3123.12 E-6 9.7
Continuation: Device-dependent Error
Error code Error text in the case of queue poll
Error explanation
241
No list defined
There is no list defined..
243
Dwell time adjusted
A dwell time given on a list cannot be processed by the unit. The setting was automatically adjusted.
251
No User Correction Table; zero assumed
An attempt has been made to switch on user correction, but no UCOR table has been stored in the
instrument yet. The instrument behaves as if a table was called which only contains 0-values.
260
Invalid keyboard input ignored
An invalid input via the keyboard is not considered.
265
This parameter is read only
An attempt has been made to change a fixedly specified value.
270
Data output aborted
Data output was aborted on the IEC/IEEE-bus.
Example: The key [LOCAL] was pressed.
304
String too long
A character string which is too long was received via the IEC bus. The names of lists may have a length
of maximally seven letters.
305
Fill pattern too long; trunctated
More data have been entered with block function FILL in the list editor than the filling range (RANGE) set
permits. The exceeding data are ignored.
306
No fill pattern specified
An attempt was made to execute a filler function without having to indicate a filler pattern.