R&S®SML / R&S®SMV03 CALibration
1090.3123.12 E-6 6.7
:CALibration:ATTenuator:STATe
The c
ommand switches ON or OFF the correction values of the attenuator.
Example: :CAL:ATT:STAT ON *RST value is ON
:CALibration:LPReset[:MEASure]?
The command calibrates Level Preset. "0" is returned for O.K. and "1" in case of an error.
Example: :CAL:LPR?
:CALibration:LFGenlevel[:MEASure]?
The command calibrates the level of the LF generator. "0" is returned for O.K. and "1" in case of an error.
Example: :CAL:LFG?
:CALibration:HARMfilter[:MEASure]?
The command calibrates the Harmonic Filters. "0" is returned for O.K. and "1" in case of an error.
Example: :CAL:HARM?
:CALibration:MULTfilter[:MEASure]?
The command calibrates the Multiplier Filters. "0" is returned for O.K. and "1" in case of an error.
Example: :CAL:MULT?
:CALibration:IFFilter[:MEASure]?
The command calibrates the IF Filters. "0" is returned for O.K. and "1" in case of an error.
Example: :CAL:IFF?
:CALibration:MAINloop[:MEASure]?
The command calibrates the Mainloop. "0" is returned for O.K. and "1" in case of an error.
Example: :CAL:MAIN?
:CALibration:FMOFfset[:MEASure]?
The command calibrates the FM offset. "0" is returned for O.K. and "1" in case of an error.
Example: :CAL:FMOF?
:CALibration:VMODulation[:MEASure]?
The command triggers a calibration for the I/Q modulator in the whole Rf frequency range. "0" is
returned for O.K. and "1" in case of an error.
Example: :SOUR:POW:ALC:TABL?:CAL:VMOD?
:CALibration[:ALL]?
The command executes automatically all calibrations in the indicated order. "0" is returned for
O.K. and "1" in case of an error.
Example: :CAL?
:CALibration:ROSCillator[:DATA]?
The command displays the calibration value entered in the Utilities - Calib - RefOsc menu.
Example: :CAL:ROSC?
:CALibration:ROSCillator:STORe
The command stores the calibration value entered in the Utilities - Calib - RefOsc menu.
Example: :CAL:ROSC:STOR