SOURce:PM R&S®SML / R&S®SMV03
1090.3123.12 E-6 6.26
[:SOURce]:PM:INTernal:FREQuency 0.1
Hz to 10 MHz
The command sets the modulation frequency.
Example: :SOUR:PM:INT:FREQ 10kHz *RST value is 1 kHz
[:SOURce]:PM:SOURce EXTernal | INTernal | TTONe
The command selects the modulation source. An external and an internal modulation source can
be specified at the same time (cf. example).
Example:
:SOUR:PM:SOUR INT
*RST value is INT
[:SOURce]:PM:STATe ON | OFF
The command switches the phase modulation on or off.
Example: SOUR:PM:STAT OFF *RST value is OFF
[:SOURce]:PM:BANDwidth STANdard | WIDE
The command sets the bandwidth for M. STANdard and WIDE are available.
Example: SOUR:PM:BAND WIDE *RST value is STAN