R&S®SML / R&S®SMV03 List of Error Messages
1090.3123.12 E-6 9.3
Command Error, continued
Error code Error text with queue poll
Explanation of error
-112
Program mnemonic too long
The header contains more than 12 characters.
-113
Undefined header
The header is not defined for the instrument.
Example: *XYZ is undefined for every instrument.
-114
Header suffix out of range
The header contains an illegal numerical suffix.
Example: SOURce3 does not exist in the instrument.
-123
Exponent too large
The absolute value of the exponent is larger than 32000.
-124
Too many digits
The number contains too many digits.
-128
Numeric data not allowed
The command contains a number which is not allowed at this position.
Example: The command SOURce:FREQuency:MODE requires the entry of a text parameter.
-131
Invalid suffix
The suffix is invalid for this instrument.
Example: nHz is not defined.
-134
Suffix too long
The suffix contains more than 12 characters.
-138
Suffix not allowed
A suffix is not allowed for this command or at this position of the command.
Example: The command *RCL does not allow for a suffix to be entered.
-141
Invalid character data
The text parameter either contains an invalid character or it is invalid for this command.
Example: spelling mistake in parameter entry; SOURce:FREQuency:MODE FIKSed.
-144
Character data too long
The text parameter contains more than 12 characters.
-148
Character data not allowed
The text parameter is not allowed for this command or at this position of the command.
Example: The command *RCL requires the entry of a number.
-158
String data not allowed
The command contains a valid character string at a position which is not allowed.
Example: A text parameter is entered in inverted commas, eg SOURce:FREQuency:MODE "FIXed"
-161
Invalid block data
The command contains errored block data.
Example: An END message was received before the expected number of data was received.
-168
Block data not allowed
The command contains valid block data at a position which is not allowed.
Example: The command *RCL requires the entry of a number.
-178
Expression data not allowed
The command contains a mathematical expression at a position which is not allowed.