R&S®SML / R&S®SMV03 Performance test
1147.7621.12 10.39 E-6
Item Instrument type Recommended
characteristics
Suitable unit R&S Order
No.
Use/measurement
9 SWR bridge 1 MHz to 3300 MHz
Directivity >40 dB
R&S ZRC 1039.9492.55/1
039.9492.52
Output reflection
coefficient
10 Modulation analyzer 100 kHz to 3300 MHz, AM, FM,
PhiM, stereo coder, stereo
decoder, distortion meter,
weighting filter ITU-R, ITU-T
R&S FMB with
option
R&S FMA-B1,
R&S FMA-B2,
R&S FMA-B3,
R&S FMA-B4
856.5005.52
855.2002.52
855.0000.52
856.0003.52
855.6008.52
Residual FM
Residual AM
AM/FM/PhiM modulation
LF generator
S
t
ereo modulation
11 Sinewave generator 10 Hz to 500 kHz,
8 V (V
peak
)
R&S ADS
R&S AFG
1012.4002.02
0377.2100.02
AM/FM/PhiM modulation
Overvoltage protection
12 AC/DC voltmeter
DC to 1 MHz R&S URE3 350.5315.03 LF generator
13 Low-noise
preamplifier
5 kHz to 3300 MHz
Gain >20 dB,
Noise figure <10 dB
Level accuracy
14 DC voltage source Setting range 0 to 10 V NGMD35 0117.7127.02 Vector modulation
15 Demodulator for
digital modulations
Error vector measurement contained in
item 2better
FSIQ 3option
K11 (for GSM)
1119.5005.03
1057.3392.02
E
rro
r vectorr
Impairments
Modulation frequency
response
16 Arbitrary waveform
generator
two channels R&S AMIQ 1110.2003.02 Vector modulation
17 Program rfor
simulation of digital
modulations
generation of data for ARB
generator
WinIQSIM,
contained in
item. 16
Vector modulation
Note:
For some settings of the R&S SMV03 the IQCW mode is mentioned. To make this mode available Lock
Level 1 has to be released. Then IQCW can be enabled by selecting the menu item
Utilities\Debug\Debugpage. IQCW can always be selected by the IEC/IEEE command
[:SOURce]:DM:IQ[:STATe] IQCW.