RF Level R&S®SML / R&S®SMV03
1090.3123.12 E-6 4.4
Attenuator Mode
Aut
o Normal setting. The electronically switched attenuator switches in
steps of 5 dB at fixed points.
IEC/IEEE-bus command :OUTP:AMOD AUTO
Fixed Level settings are made without switching the attenuator (see
section "Non-Interrupting Level Setting").
IEC/IEEE-bus command :OUTP:AMOD FIX
Atten Fixed Range
Indicates the level range of non-interrupting level setting in "Attenuator Mode
Fixed".
IEC/IEEE-bus command :OUTP:AFIX:RANG:LOW?
(only lower value querable)
Knob Step User
Input value of step size of level variation via rotary knob. The RF level is varied
by the entered step size if Knob Step is set to User.
IEC/IEEE-bus command :SOUR:POW:STEP 1
Knob Step
Decimal The variation step size corresponds to the position of the digit cursor.
User User-defined, the variation step size is as entered under Knob Step
U
s
er (only in dB).
IEC/IEEE-bus command :SOUR:POW UP / :SOUR:POW DOWN
(The RF level is always in-/decremented by the
given step value, no special switching to Knob
Step User.)
Power Resolution
Selection of resolution of level display
0.1 dB The resolution of the level display is 0.1 dB.
0.01 dB The resolution of the level display is 0.01 dB.
Power On State
Selection of status to be assumed by RF output after power-up of the
instrument.
RF Off The RF output is switched off.
Previous Setting The RF output assumes the status active before switch-off.
IEC/IEEE-bus command :O
UT
P:PON OFF
Exclude from Recall
Off Normal setting. The stored RF level is loaded too when instrument
settings are loaded with the [RCL] key.
IEC/IEEE-bus command :SOUR:POW:RCL INCL
On The stored RF level is not loaded when instrument settings are
loaded, ie the current level setting is maintained.
IEC/IEEE-bus command :SOUR:POW:RCL EXCL
Preset RF State
Off After preset, the RF state is "Off",
On After preset, the RF state is "Off".
IEC/IEEE-bus command -

Table of Contents

Related product manuals