Reference Manual
00809-0100-4026, Rev FA
July 2009
7-19
Rosemount 5400 Series
Figure 7-16. Holding and Input
Registers can be viewed in
RRM.
RESET TO FACTORY
SETTINGS
This function resets all, or a specific part, of the holding registers to the factory
settings. It is recommended that a backup of the configuration be made
before resetting, so the old transmitter configuration can be loaded, if
necessary.
RRM: choose menu option Tools > Factory Settings.
Figure 7-17. The Reset to
Factory Settings window in
RRM.
AMS Suite: Tools / Service > Factory Settings.
HART Command: [1, 2, 8].