Reference Manual
00809-0100-4026, Rev FA
July 2009
Rosemount 5400 Series
7-20
SURFACE SEARCH The Surface Search command triggers a search for the product surface and
can be used, for example, if the measured level is locked onto a disturbing
object in the tank (see “Configuration Report” on page 7-18).
USING THE SIMULATION
MODE
This function can be used to simulate measurements and alarms.
RRM: choose menu option Tools > Simulation Mode:
Figure 7-18. The Simulation
Mode window in RRM.
AMS Suite: Tools > Service > Simulation Mode.
HART Command: [1, 2, 2].
WRITE PROTECTING A
TRANSMITTER
A Rosemount 5400 Series transmitter can be password protected from
unintentional configuration changes. The default password is 12345 and it is
recommended that this password not be changed to facilitate service and
maintenance of the transmitter.
RRM: Tools > Lock / Unlock Configuration Area.
AMS Suite: Tools > Service > Lock / Unlock Device.
HART Command [1, 2, 7].