Reference Manual
00809-0100-4026, Rev FA
July 2009
Rosemount 5400 Series
B-10
National Supervision and
Inspection Center for
Explosion
Protection and Safety of
Instrumentation (NEPSI)
Approvals
GYJ06242X, GYJ06458X
Figure B-8. Approval Label
National Supervision and
Inspection Center for Explosion
Protection and Safety of
Instrumentation (NEPSI)
E3 Flameproof:
Ex iad IIC T4 (-40°C<T
a
<+70°C
(1)
).
U
m
=250 V
I3 Intrinsically Safe:
Ex ia IIC T4 (-40°C<T
a
<+70°C
(1)
).
4-20 mA / HART
®
model: U
i
=30 Vdc, I
i
=130 mA, P
i
=1.0 W,
C
i
=7.26 nF, L
i
=0 H.
FISCO model: U
i
=17.5 Vdc, I
i
=380 mA, P
i
=5.32 W, L
i
=C
i
=0.
Installation Drawing: 9150079-907.
(1) +60°C with FISCO option.