IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001 205
Rise-time Control Function
Yes
RAMP (Ramp Time)
0.1~999.9S
TIMER (Test Time)
OFF
1
, 0.5S~999.9S
GND
ON/OFF
1
The timer can only be turned off under special MANU mode (MANU=***-000).
Insulation Resistance Test
Output Voltage
50V~1000V *IIT-5010 also includes a 125V test point.
Output Voltage Resolution
50V
Output Voltage Accuracy
(1% of setting+5V) with no load
Resistance Measurement
Range
1M~ 9500M (IIT-2010)
1M~ 50G (IIT-5010)
Test Voltage
Measurement
Range
Accuracy
(IIT-2010)
50V≤V≤450V
1~50MΩ
±(5% of reading +1 count)
51~2000MΩ
±(10% of reading +1 count)
500V≤V≤1000V
1~500MΩ
±(5% of reading +1 count)
501~9500MΩ
±(10% of reading +1 count)
(IIT-5010)
50V≤V≤450V
0.001~0.050G
±(5% of reading +1 count)
0.051~2.000G
±(10% of reading +1 count)
500V≤V≤1000V
0.001~0.500G
±(5% of reading +1 count)
0.501~9.999G
±(10% of reading +1 count)
10.00~50.00G
±(20% of reading +1 count)
Output Impedance
600kΩ
Window Comparator Method
Yes