IIT-2000/2010/5000/5010 Instruction Manual/English
204 05/08/2015 Version No. 001
Current Best Resolution
IIT-2000/2010:
1uA
0.001mA(0.001mA~0.999mA)
0.01mA(01.00mA~09.99mA)
0.1mA(010.0mA)
IIT-5010/5000:
1uA
0.001mA(0.001mA~1.100mA)
0.01mA(01.11mA~11.00mA)
0.1mA(011.0mA~020.0mA)
Current Measurement
Accuracy
IIT-2000/2010:
± (1.5% of rdg + 30 counts) when HI SET <1.00mA
(1.5% of rdg + 3 counts) when HI SET ≥1.00mA
IIT-5010/5000:
± (1.5% of rdg + 30 counts) when HI SET <1.11mA
(1.5% of rdg + 3 counts) when HI SET ≥1.11mA
Judgment Valid Range
(DCW)
HI SET:
0.011mA to 1.100mA
00.11mA to 10.00mA(IIT-5010/5000 11.00mA)
001.1mA to 020.0mA(IIT-5010/5000 only)
LOW SET:
0.010mA to 1.099mA
00.10mA to 09.99mA(IIT-5010/5000 10.99mA)
001.0mA to 019.9mA(IIT-5010/5000 only)
Window Comparator Method
Yes
ARC DETECT
Yes

Table of Contents