IIT-2000/2010/5000/5010 Instruction Manual/English
05/08/2015 Version No. 001 51
ACW (LO)
DCW (LO)
IR (HI)
GB (LO)
0.000mA~041.9mA (IIT-2000/2010)
0.000mA~109.9mA (IIT-5000/5010)
0.000mA~010.9mA (IIT-2000/2010)
0.000mA~020.9mA (IIT-5000/5010)
0001MΩ~9999MΩ, ∞ (IIT-2010)
0.001GΩ~50.00GΩ, ∞ (IIT-5010)
000.0mΩ ~ 649.9mΩ
*Please note that the resolution of the measured value
depends on the resolution of HI SET setting.
The LO SET setting is limited by the HI SET setting. The LO
SET limit cannot be greater than the HI SET limit.
When setting the current, be aware that a maximum of 200VA
can be set for ACW and 50W for DCW (IIT-2000/2010) or
500VA and 100W, respectively for IIT-5010/5000.