IIT-2000/2010/5000/5010 Instruction Manual/English
78 05/08/2015 Version No. 001
Pressing the STOP button will return the tester
to the READY status.
Pressing the START button will restart the test.
The buzzer will only sound if the Pass Sound is set to ON.
See page 113 for details.
The START button is disabled when the buzzer is beeping.
The timing diagrams below show the ACW, DCW, IR and GB
timing for the START status, TEST status and PASS judgment.
Discharge time
Output V
time
Initial time
START
TEST
PASS
RAMP TEST TIME
(Approximately 150ms)