Safety Precautions
(h)
WARNING
Unplug the power cord from the power outlet and store it while the instrument is not in use.
For safety, if the instrument will not be used for an extended period of time, clean each part
and store the instrument in a place free from humidity and from dust and foreign matter.
Using the instrument in a humid place, place exposed to oil smoke or steam, or dusty place, may
cause its parts to deteriorate, short-circuit, fire, etc. The installation humidity range is 30 to 85% (non-
condensing). If you wish to install the instrument in a special place, consult the Sakura Finetek
Technical Support representative or local representative.
If dews are formed on the instrument, do not turn on the power.
If the power is turned on while dews are formed on the instrument, current leak may occur and cause
fire or electric shock. Do not connect the instrument to a power-supply facility until dews have
evaporated and the instrument has become completely dry.
Do not operate the power plug and switches with a wet hand. When connecting/ removing the
power plug, do so by holding the plug.
Failure to observe these instructions may result in electric shock.
Be sure to use the supplied power cord. Also, be sure to use a correct power cord that is
suitable for the current capacity of region where the instrument is used.
If an inappropriate power cord or voltage is used, instrument malfunction due to short-circuit, fire or
voltage drop, unexpected accidents or injuries due to electric shock, etc., can occur. The preset
voltage cannot be changed by the user.
For the power cord, a standard cord specified by local law may have to be used.
If a power cord not specified by the PSE standard, UL/CSA standard or EC standard is required,
contact the Sakura Finetek Technical Support representative or local representative.
Do not forcibly bend, pull, twist or bundle the power cord and cables or otherwise damage the
power cord and cables.
Fire or electric shock may occur.
Do not overload the outlet and do not use an extension cord.
Doing so may stop the operation of the instrument due to voltage drop, or fire can occur due to heat
generation.
This instrument has been designed and tested according to CISPR11 Class A. When the
instrument is used in a home, it may cause electromagnetic interference, in which case
measures to reduce interference may have to be taken.
Radiant waves may prevent this instrument from operating properly, so do not use the
instrument near a radiation source of strong electromagnetic waves.
Use the instrument under the conditions (such as voltage within 10% of the rated voltage)
stated in the Specification.

Table of Contents

Other manuals for Sakura Tissue-Tek Prisma Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Prisma Plus and is the answer not in the manual?

Sakura Tissue-Tek Prisma Plus Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Prisma Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals