Safety Precautions
(g)
WARNING
A wrong operation may stop or interrupt the tissue processing.
Be careful not to let unauthorized personnel manipulate or tamper with the instrument while
operating.
A wrong operation may stop or interrupt the tissue processing. (If the power is turned off by
unauthorized personnel, the stain process may be interrupted and the staining problems may occur.
Arrange the work environment so that only the operators with expert knowledge can use the
instrument. (Also consider using external alarms, etc., just in case.)
Do not bring fire sources close to the instrument.
The instrument uses organic solvents that may ignite upon contact with fire.
Do not let foreign matter go inside the instrument.
If water, metal, paper or other foreign matter enters the air inlets (to the fan, etc.), fire, electric shock,
injury of the user, instrument failure, etc., may occur. If foreign matter entered the instrument,
immediately stop the instrument and contact the Sakura Finetek Technical Support representative or
local representative.
Do not operate the instrument while wearing metal accessories.
Conductors (metal pen, accessories, etc.) worn on your body may contact the instrument and cause
electric shock.
Do not block the intake/exhaust ports and air inlet around the instrument.
Do not block the intake/exhaust ports (air inlets) with a wall or object or use the instrument in a poorly
ventilated place or with the dustproof covers still attached, because heat will be trapped in the
instrument and cause the capacity of the instrument to drop, while fire, accident, failure, etc., may also
occur due to abnormal overheating.
Periodically check the power cord and cables and power outlet for damage and accumulation
of dust.
Foreign particle and dust accumulating on the power outlet may cause fire. Clean the power outlet
periodically.
Do not let tissue with flammable solvent attached go inside the drying station.
Deposited flammable solvent may be heated by the drying station and vaporize, causing fire or
explosion.
Do not install the instrument in a place used as a living area.
If a common area is used, an exhaust system is required. Provide electrostatic discharge protection
in the installation location.
Install the instrument indoors. Do not place the instrument in a location exposed to direct
sunlight, snow or rain.
This product is specified for indoor use, so do not use it outdoors. Ultraviolet rays and high
temperature may damage the instrument. Never use the instrument while exposing it to snow or rain
because it may cause electric shock or short-circuit.

Table of Contents

Other manuals for Sakura Tissue-Tek Prisma Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek Prisma Plus and is the answer not in the manual?

Sakura Tissue-Tek Prisma Plus Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek Prisma Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals