Commissioning
PR 5220 Instrument Manual
EN-50 Sartorius
4.3.8.1 Weighingpoint ‚Internal A’
- Calibration
Calibration of weighing electronics
- New
Reset Span and deadload
Contin, Cancel
- Max
0.00001 ...<3000>... 999999 <kg>, t, lb, g
- Scale interval
<1>, 2, 5, 10, 20, 50
- Deadload at
<0.000000 mV/V> or [by load]
- Max at
<1.000000 mV/V> or [by load]
by load
0.00001 ... 999999 <kg>, t, lb, g
- Calibrated at
(Display only)
- Sensitivity (µV/d)
(Display only)
- Test
Determine test value
Exit calibration
Save or cancel changes
- Modify
Only for minor modifications/ setting new dead load, otherwise [New
see New
- Param
- Measuretime
5 ms, 10, 20, 40, 80, 160, <320>, 640, 960, 1280, 1600ms
- Digital filter
<off>, Bessel, aperiod., butterw., tscheby.
- External supply
<8 -12 VDC>, below 8 VDC
- * Fcut
Cut-off frequency, only unless filter 'off', 0.1 - 80.0 Hz
- Test mode
<Absolute>, relative
- W & M
<none>, OIML, NSC, NTEP
- Standstill time
0.01 s…<0.50 s> ... 2.0 s (range is dependent on response time)
- Standstill range
0.00 d ... <1.00 d> ... 10.00 d
- Tare timeout
0.1 s ... <2.5 s> ... 25 s, timeout due to instability
- Zeroset range
0.00 d ... <1.00 d> ... 10.00 d
- Zeroset range
0.00 d ... <1.00 d> ... 10.00 d
- Zerotrack step
0.00 d ... <1.00 d> ... 10.00 d
- Zerotrack time
<0.0 s> ... 25 s
- Overload
0 d ... <9 d> ... 999900 d
- Min
0 d ... <50 d> ... 999900 d, minimum weight
- Range mode
<Single range>, multiple range, multi-interval
- * Range limit 1
In weight, unit same as Max, transition from small to medium scale
interval, *only for multiple range or multi-interval
- * Range limit 2
In weight, unit same as Max, transition from medium to high scale
interval, *only for multiple range or multi-interval
- View(when - View(when CAL switch closed)
- Max
(Display only)
- Scale interval
(Display only)
- Deadload at
(Display only)
- Max at
(Display only)
- Calibrated at
(Display only)
- Sensitivity (µV/d)
(Display only)
- Param
Items as for Param. (display only)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius PR 5220 Series and is the answer not in the manual?

Sartorius PR 5220 Series Specifications

General IconGeneral
BrandSartorius
ModelPR 5220 Series
CategoryTransmitter
LanguageEnglish

Related product manuals