PR 5610 Installation Manual Commissioning
Sartorius 97
[Profibus-DP]
Enter value for the Profibus-DP address with numeric keys
[1…126] and confirm with .
[DeviceNet]
Select baud rate [125k], [250k] or [500k] with or ;
confirm with
.
Enter value fort he DeviceNet address [1 ... 62] with numerical
keys; confirm with .
For all protocols the [Scale- interface] has to be set.
Select the operating mode [enabled] for a standard interface
fieldbus - PR 5610 with 8 Bytes and [disabled], if a IEC 61131-
application program is handling the fieldbus; confirm with
.
Leaving the menu
Leaving the fieldbus configuration menu.
Select [YES] for saving the fieldbus configuration, the instru-
ment will perform a WARM start.
Select [NO] for going back to [Setup]-[Fieldbus Parameter].
After saving the fieldbus configuration, the instrument makes a warm start in order to initialize the PR 1721
module.
4.8.5 Licence setup
Additional application programs e.g. licence 102 FLOW and general software functions e.g. PR 1713/21 can be
activated via a licence setup menu.
The access is carried out via the menu[Setup]-[Licence Setup].
Select the access with .
Select the function [Show], [Add] or [Delete].
After [Show] an overview of the function licences can be
displayed with
or . The function licences are displayed
one after another with their status [Enabled] or [Disabled].
After that entered licences for application programs can be
displayed with
or . Leave the menu with .
After [Add] a 7 digit licence number can be entered and con-
firmed with . For a valid number the message 'PR17XX/XX
enabled' is displayed. Otherwise the message 'wrong Licence
number' is displayed. Leave the menu with
.