Commissioning PR 5610 Installation Manual
60 Sartorius
4.2.3 Operating concept ‘Commissioning’
The instrument configuration is done via keypad or, alternatively, with a terminal or a terminal program on a
PC. The sections for calibration and configuration can be reached from the main menu via [Setup]. Gross weight
display for calibration, if necessary, please: reset tare + set zero!
Operating concept: The instrument is menu-guided with softkeys via keypad. Inscriptions on the 3 softkeys
are with up to 6 characters per key on the lower display line. The symbols in the lower line of the alphanumeric
display limit the softkey inscriptions. Thereby,
means: only these softkeys, and means: further functions are
available with
. In the description of operation procedures with softkey functions the softkey symbol is not
indicated, but the function (text) itself is indicated bold in square brackets.
Example:
left softkey = [YES], right softkey = [NO]
During editing,
must be used also for selecting the weight unit [g , kg , t , lb]
Example:
change FSD with numeric keys
and unit with
Depending on function, the input modes are switched over between selection from a menu / a selection box or
entry of alphanumeric characters (editing). The relevant value is displayed beside the parameter, i.e. a quick
survey of all configuration settings is possible by parameter 'scrolling'. For editing e.g. parameter [Baudrate], it
must be selected with
.
The parameters can be left either by
or . In field [Save], after changes, answer request for saving:
Example:
[YES]: changed values are saved
[NO]: no change, back to last menu