PR 5610 Installation Manual Commissioning with Terminal
Sartorius 111
5.2 Configuration
Calling up mask
Configuration
is from the
Main-Command-Level
, by pressing key [C].
PR1713 Rel. 03.12. 20.05.2004 09:05
Configuration Mo 05.July 2004 13:06:16
D Set date
T Set time
C Change serial port parameter
L Add, delete, show license
S Set software configuration
U Set units
F Set fieldbus configuration
P Print all configuration data
E Exit
If a production is active, configuration is disabled:
Production is active, no configuration possible. Hit any key to continue
Press any key to remove this message from the display.
Function survey:
Set date
D Set date
Set time
T Set time
Change serial port parameters
C Change serial port parameter
Add, delete and display licences
L Add, delete, show license
Change and display software configuration
S Set software configuration
Entry of user-defined units
U Set units
Parameter of fieldbusoption PR 1721/0x
F Set fieldbus configuration
Print out all configuration data
P Print all configuration data
Leave menu Configuration
E Exit
5.2.1 Set date
PR1713 Rel. 03.12 20.05.2004 09:05
Configuration Mo 05.July 2004 13:06:16
D Set date
T Set time
......
Enter new date [30.11.2001]:
Press key [D]
Message:
Enter new date [30.11.2001]: _
is displayed
Enter the new date e.g.:
02.12.2001
Press key [], the line is removed from the display if the entry was correct. With faulty entry,
message:
(bad date) enter new date [30.11.2001]: _
is displayed.
Check against the header (top right)