PR 5610 Installation Manual Commissioning
Sartorius 73
4.5.4 Select weighing point
The selection of the weighing point is carried out by [Setup]-[Weighingpoints].
Change weighing point with or and select calibration of
the displayed weighing point with [Calib].
The calibration menu with the selections [New], [Modify] or
[Param] is displayed.
4.5.5 Display calibration data
Display further functions with .
Select [View].
All calibration parameters and data are displayed as these have been entered / calculated during calibration
procedure.
4.5.6 Select calibration mode
[New]
[New] starts a calibration with default values.
Safety prompt; if [YES] default values are loaded and calibration
procedure starts.
In all display clips the default value is displayed. Default values for [New]:
Fullscale
Stepwidth
Deadload
Span
[Param]
<3000> <kg>
<1>
<0.000000 mV/V>
<1.000000 mV/V>
All parameters remain unchanged !
[Modify]
This menu should only be used for small changes, in other cases [New] has to be selected.
[Modify] changes current calibration data (fullscale, weight
unit, stepwidth, deadload and span).