PR 5610 Installation Manual Hardware Configuration
Sartorius 25
3.3.3 Bar code reader
The DIN socket (at the rear side of the housing) can be used to connect a bar code reader instead of an external
keyboard.
Please check the power consumption of the bar code reader before connecting,
it may not exceed 50 mA @ 5 V.
Overload is prevented by a multifuse element.
For PIN layout of the DIN socket refer to chapter 3.3.2.
At the instrument the keyboard character set has to be set to US (default). If a GERMAN keyboard had been
used, return to US with [Strg][F1] before connecting the reader.
Bar code reader setup (Table: Opticon OPL 6735 wedge) is done by scanning the respective codes of the
handbook.
Default values are indicated with (d).
Opticon CPL 6735 wedge
Linker: AT wedge
Keyboard layout: with keyboard (d)
Keyboard language: US (d)
Intercharacter delay for
wedges:
Delay = 10
Set suffix: Clear all suffixes
Read mode options: Single read
Redundancy: 3 times
redundant
Further bar code readers to be used are: Syntech (Cipher 1021G) and Gryphomn (M-100).
For safe readings it is recommended to set the redundancy to 3 readings.
To acknowlege each reading via keypad on the instrument, the auto carriage return function of the bar code
reader has to be switched off.
CAUTION:
With bar code reader connected, protection class IP 65 is not met any more!