EasyManua.ls Logo

Sartorius PR 5610 - PR 1713;17 6 Opto Inputs; 8 Opto Outputs

Sartorius PR 5610
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Hardware Configuration PR 5610 Installation Manual
52 Sartorius
3.4.10 PR 1713/17 6 Opto inputs / 8 Opto outputs
Max. 3 cards PR 1713/17 can be used. The digital interfaces are passive (external supply is necessary) and have
one common potential per input group and one common potential per output group.
Connecting method: 2-row 25-pin connector
Terminal block A/B with 2x8
screw terminals
Number of inputs / outputs:
Inputs: 6, outputs: 8
Input signal:
Low: 0 ... 4 V or open
High: 10 ... 31 V
External supply required
Input current: < 7 mA at 24 V
< 3 mA at 12 V
Output: Max. switching voltage: 31 V
Max. switching current : 25 mA
Voltage drop at 25 mA: 3 V
External supply required
Potential isolation: Yes, 6 inputs have got 1 common
minus potential, 8 outputs have
got 1 common minus potential
Dimension (LxWxH): 80 x 68 x 15 mm
Weight: 33 grs.
Pin allocation of option connectors in the unit (installation see chapter 3.2.1)
Terminal block B input B1 B2 B3 B4 B5 B6 B7 B8
upper row 6 channels + CH1 + CH2 + CH3 + CH4 + CH5 + CH6 - INP - OUT
Terminal block A output A1 A2 A3 A4 A5 A6 A7 A8
lower row 8 channels + CH1 + CH2 + CH3 + CH4 + CH5 + CH6 + CH7 + CH8