4.4.1 Error messages on the weight display
The internal ADC and the external weighing points can generate error messages to appear on the weight
display. If more than 1 weighing point has been assigned, the selection is done with . The messages are
displayed with 'Error X'
.
Error messages on the weight display
Error 1 internal arithmetic overflow (faulty calibration values)
Error 2 input voltage is above FSD + overload
Error 3 input voltage is above the permissible range of 36 mV. However, an error in the analog section,
a defective load cell, or a cable break are also possible.
Error 4 weight value exceeds display digits
Error 5 weight is not available, e.g. weighing point is busy
Error 7 input voltage is negative or incorrect load cell connection
Error 8 ADC error, e.g. internal ADC defective or overloaded
Error 9 no communication with weighing point
Error 11 weight is not available
Error 15 serial number check failed