Technical Data PR 5610 Installation Manual
156 Sartorius
6.5.2 GSD file for PR 1721 Profibus DP
Use our website at Internet address http://www.global-weighing.com for loading the GSD file GWT_00C3.GSD for
PR 1721, or generate file GWT_00C3.GSD by entry of the following listing (the comments behind the semi-colon
are optional):
;=======================================================================
;GSD-Datei für das Profibus-Interface von PR1721/01
;Stand 21.08.2000 - GLOBAL Weighing Technologies Version:V1.1
;=======================================================================
;
#Profibus_DP
Vendor_Name = "GLOBAL Weighing Technologies"
Model_Name = "PR1721/01 Profibus Interface"
Revision = "Version 1.1"
Ident_Number = 0x00C3
Protocol_Ident = 0 ;Nur DP-Gerät
Station_Type = 0 ;Kompakt Station
FMS_supp = 0 ;FMS wird nicht unterstützt
Hardware_Release = "1.0"
Software_Release = "1.0"
;
9.6_supp = 1 ;unterstützte Baudraten
19.2_supp = 1
93.75_supp = 1
187.5_supp = 1
500_supp = 1
1.5M_supp = 1
;
MaxTsdr_9.6 = 60 ;max. Antwortzeiten bei den unterschiedlichen Baudraten
MaxTsdr_19.2 = 60
MaxTsdr_93.75 = 60
MaxTsdr_187.5 = 60
MaxTsdr_500 = 100
MaxTsdr_1.5M = 150
Redundancy = 0 ;keine redundante Übertragung
Repeater_Ctrl_Sig = 0 ;kein RTS Signal
24V_Pins = 0 ;keine 24V für den Anschluß eines Wartungsgerätes
Freeze_Mode_supp = 1 ;Freeze Modus wird unterstützt
Sync_Mode_supp = 1 ;Sync Modus wird unterstützt
Auto_Baud_supp = 1
Set_Slave_Add_supp = 0
User_Prm_Data_Len = 0
Min_Slave_Intervall = 50 ; 50x100uS->5ms. min. Slave intervall
Modular_Station = 0
Module = "8 Byte Out, 8 Byte In" 0xA7, 0x97
EndModule