PR 5610 Installation Manual Commissioning
Sartorius 65
4.3.2 Boot menu
The Boot menu can be accessed by the following actions:
Pressing during instrument power up
Pressing + , if [Setup]-[Software Parameter]-[Reset on stop+exit] = 1 or 5 s,
a [Warm] start is not possible!
Menu [Setup]-[Reboot]
• If a warm start is possible:
[Cold] All programs and tables loaded in RAM are deleted.
[Warm] All data remain unchanged.
[Flash] Program loading (also parts).
[Test] Can be selected with .
• If a warm start is not possible:
[Cold] All programs and tables loaded in RAM are deleted.
[Erase] Erases the EAROM (only with CAL-switch = open).
[Flash] Program loading (also parts).
[Test] Can be selected with .
With [Erase] all calibration and configuration data are adjusted to the default (factory) settings.
Possible error message:
If CAL-switch is closed.
Safety prompt:
[YES] Data are overwritten.
[NO] All data remain unchanged.
• Warm start possible (with Reboot)
If [Bios] is selected, a [Warm] start is not possible
anymore!
[Test] can be selected with .