PR 5610 Installation Manual Commissioning
Sartorius 91
The previously stored user weight is shown, the weight
can be altered and confirmed with .
The target weight for calibration is shown, the calibration
weight has to be placed on the scale.
The deviation is shown. After selecting [Accept] the
instrument returns to menu [xBPI Scale adjust].
If the parameter [Confirming adjust.] has been set to
[automatically], no confirmation is necessary.
4.7.10 xBPI Calibration with automatic weight recognition
The procedure is the same as in chapter 4.7.8 xBPI Calibration with internal weight. Instead of [Intern] the
function [Auto] has to be selected.
4.7.11 xBPI Calibration with preset weight
The procedure is the same as in chapter 4.7.8 xBPI Calibration with internal weight. Instead of [Intern] the
function [Deflt.] has to be selected.