PR 5610 Installation Manual Commissioning
Sartorius 61
4.3 Main menu
After connection to the power supply, the instrument shows the version number "PR 1713 Rel. 02.00.0" or the
name of the application package (e.g. IBC Controller).
The unit can be configured via the keypad or (after switching over) on a terminal. The operation of the unit is
menu-guided. Calibration and configuration are accessible via menu [Setup].
Example:
Depending on the application package certain
functions can be selected
Further items are displayed with
Menue items depend on the application package (e.g. Batch-X5).
Without loaded applications only [Setup] is displayed.
Overview on the menu items in the main menu
Menu item Description
[Setup] Configuration, calibration, interface definition, initial data
[Atest] Display of the testvalue for the internal ADC, which was calculated during calibration
'other' Depending on application package, see respective operating manual
[Setup]-[Config] Configuration for application program, see respective operating manual
4.3.1 Function setup
In menu [Setup] the most important parameters can be set. Configuration of the instrument depends on the
application package, licences and the built-in options.
Calibration of the internal WP is done in a dialogue. During calibration, all parameters which can be entered are
checked. Storage of values which do not make sense is not possible. However, compliance with the relevant
(Legal Metrology) standards must be checked by the installer (calibration officer). For protecting calibration
data from overwriting, the CAL-switch must be closed (disconnect the instrument from power supply, open it,
close the CAL-switch (see chapter 3.3), close the instrument and connect it to the power supply).
Configuration can be done without additional terminal via front panel display and keypad. Simply use the
scrollkeys and the softkeys in connection with the alphanumeric display. Within setup menu, all items in each
level are scrolled in an infinite loop, which can be left to the next higher level by pressing
. Exception: the
calibration loop must be terminated by the function keys [Save], [Undo]. A displayed or selected information is
entered by pressing
key.