PR 5610 Installation Manual Commissioning
Sartorius 81
Exit configuration and store data with [Yes]. With [No] the data
remain unchanged.
Automatic return to the weighing point selection menu.
4.6.2 Weighing point function GWT-scale
With the GWT-scale the update time must be defined and with fieldbus applications additionally a slave address
must be defined.
With [Config] the configuration of the GWT-scale can be
accessed.
Select W&M mode [none], [OIML], [NTEP] with or ; confirm
with
.
Enter the overload range > FSD with numeric keys from 0 to
9999999 d; confirm with .
Enter minimum weight for print-out with numeric keys from 0 to
9999999 d; confirm with .
Enter time interval for weight data transfer from the external
weighing point with numeric keys; confirm with .
Enter slave number for fieldbus application with numeric keys;
confirm with .
Enter serial number of a connected scale with numeric keys, if it
has to be checked ; confirm with
. In the case of non-
conformity error 15 is displayed.
Automatic return to the weighing point selection menu.
4.6.3 Weighing point function User-scale
The user defined weighing point function delivers weight data via a SPM variable.
With [Config] the configuration of the user-scale can be accessed.
Select W&M mode [none], [OIML], [NTEP] with or ; confirm
with
.
Enter numeric value for FSD with or without decimal point.
Change weight unit with
; confirm with .
Select the step width 1, 2, 5, 10, 20 or 50 with or and
confirm with
.
Enter the overload range > FSD with numeric keys from 0 to
9999999 d; confirm with
.