Commissioning PR 5610 Installation Manual
84 Sartorius
4.7.1 xBPI Serial port setup
In menu [Setup]-[Serial Ports]-[Serial port setup] the port connected to the Sartorius scale / platform has to be
defined.
Select with or the interface module and confirm
with
.
If currently a protocol is running, it will be stopped and
shown.
With protocol can be selected.
Select with or the xBPI protocol
and confirm with
.
If necessary, select and alter further parameter with or
according to the configuration of the xBPI device.
The xBPI protocol is started.
Leave the serial port setup with
.
4.7.2 xBPI Weighing point assignment and configuration
In menu [Setup]-[Weighingpoints]-[WP X] the scale can be assigned to the logical WP A or B and configurated.
Select [Weighingpoints], confirm with .
Select with or the logical WP B, select with [Assign].
Select with or [xBPIScale] and confirm with .
Select [YES] for assignment of xBPI scale to WP B.
Select [Config].
For configuration refer to chapter 4.6