4.1.2 Configuration data
Adjusting the serial interface/ communication, field bus parameters. The configuration settings are also saved by
command. The configuration settings are independent of CAL-switch, i.e. access to the configuration settings is
always possible.
5. If the EAROM is erased, the default configuration data are loaded into RAM (working memory).
6. During configuration, the configuration data in RAM are modified.
7. Function [Save software conf ?]-[YES] stores the modified configuration data from RAM into EAROM.
8. At power-on, the configuration data are loaded from EAROM into RAM.
4.1.3 Programming with PR 1750
Programming with PR 1750 is only possible with opened CAL-switch. Project files (P32) can be loaded on-line
directly by EW-Com.
Firmware (.hex) can be loaded off-line with menu [Flash], see chapter 4.3.2.2. FlashIt!32 also can merge project
files.
4.1.4 Cold / Warm start
With
kept pressed during switch-on (at power recovery), the boot menu is displayed:
The user can select:
[Cold]: Cold start - program and database are erased, interface configuration and calibration data remain
unchanged, options in slot 1-4 are detected
[Warm]: Warm start - all data remain unchanged
In a further menu item, the default settings can be reestablished. Thereby, calibration data, interface
configuration and hardware clock are set to standard values.
The boot menu can be called up from [Setup]-[Reboot]:
If [Bios] is selected, the booting is started, setting the instrument to default data.