Commissioning PR 5610 Installation Manual
78 Sartorius
FSD is smaller than the calibrated weight (span).
4.5.13 Set parameters
[Param]
Select [Param] to enter the measuring parameters.
Enter measure time with numeric keys from 50 to 1000 ms;
confirm with .
Select choices [on] or [off] for analog filter for the measuring
signal with or ; confirm with . See chapter 5.5.2.
Select digital filter characteristics [off], [bessel], [aperiod.],
[butterw.] or [tscheby.] with
or ; confirm with . See
chapter 5.5.2.
Enter cut-off frequency with numeric keys from 0.1 to 8.0 Hz;
confirm with .
Select mode for the test value [Absolute] or [Relative] with or
; confirm with .
Select W&M mode [none], [OIML], [NTEP] with or ; confirm
with
.
Select load cell excitation voltage [12 Volt] or [20 Volt] with
or
; confirm with .
Enter standstill interval with numeric keys from 0.01 to 1.6 s;
confirm with .
Enter standstill tolerance with numeric keys from 0.01 to 10 d;
confirm with .
Enter time for aborting an unexecutable tare / set zero command
with numeric keys from 0.1 to 25 s; confirm with .
Enter zero set range with numeric keys from 0.01 to 500 d;
confirm with .
Enter zero track range with numeric keys from 0.01 to 500 d;
confirm with .
Enter zero track step with numeric keys from 0.01 to 10 d;
confirm with .
Enter zero track time interval for the automatic zero setting with
numeric keys from 0.0 to 25 s; confirm with .