PR 5610 Installation Manual Commissioning
Sartorius 79
Enter the overload range > FSD with numeric keys from 0 to
9999999 d; confirm with .
Enter minimum weight for print-out with numeric keys from 0 to
9999999 d; confirm with .
Select operating mode multi range scale (max. 3 ranges) [on] or
[off] with
or ; confirm with . The entry of the switch
over points appears only for selection [on].
Enter switch over point from range 1 to range 2 from 0 to
9999999 d (0 = only 1 range) with numeric keys; confirm with .
Enter switch over point from range 2 to range 3 from 0 to
9999999 d (0 = only 2 ranges) with numeric keys; confirm
with
.
Exit parameter entry mode with . Automatic jump to the
selection menu for the calibration with the indication, that the
calibration has been changed. Continue with 4.5.12 Exit
calibration.
4.5.13.1 W&M mode
In menu [Setup]-[Weighingpoints]-[Calib]-[Param]-[W&M] the options are possible: [none], [OIML], [NTEP] and
[NSC].
The main difference is the appearance of the weight unit on the display and the signal range:
weight none OIML NTEP NSC
negative unit no unit unit no unit
> FSD no unit no unit unit no unit
> FSD + overload ERROR 2 ERROR 2 ERROR 2 ERROR 2
gross indication B B G G
> 0.03 mv/V at
3000 d
> 0.3 mv/V at
3000 e
> 0.3 mv/V at
3000 e
> 0.25 mv/V at
3000 e
min.
measuring signal
> 0.05 mv/V at
5000 d
> 0.5 mv/V at
5000 e
> 0.5 mv/V at
5000 e
> 0.42 mv/V at
5000 e
4.5.13.2 Multirange scale
The multirange function is controlled by three parameters in menu [Setup]-[Weighingpoints]-[Calib]-[Param].
With [Multirange mode] = [on], the scale has 3 ranges with 3 different resolutions: The change-over limits
[Multirange limit 1] and [Multirange limit 2] represent the range borders. If the gross weight is larger than
range 1, the next higher range will be selected and the next higher stepwidth is valid (1->2->5->10->20->50).
If the gross weight is <= 0.25 d of range 1, the scale in standstill and untared condition, the scale will be
switched back to range 1. On the display the ranges are marked by quotes or dashes ( ' range 1, " range 2, =
range 3 ) left of the weight value.
During calibration the multirange function is principally switched off.
For 'legal for trade' application a label has to be placed stating Max., Min. and e per range.