PR 5610 Installation Manual Commissioning
Sartorius 77
Automatic return to the selection menu for the calibration with
the indication, that the calibration has been changed.
4.5.12 Exit calibration
Finish the calibration with .
Display of the choices [Save] and [Undo].
[Save]
Store new calibration data with [Save].
Acknowledgement of leaving the calibration menu.
Automatic jump in the selection menu for weighing points.
[Undo]
Leave the calibration menu without data storage with [Undo].
Automatic return to the selection menu for weighing points.
If was pressed during calibration procedure:
Safety prompt, if
has been pressed during calibration. With
[NO] the instrument remains in the calibration menu.
With [YES] the calibration will be finished. Automatic jump in the
calibration menu.
4.5.12.1 Calibration error messages
FSD cannot be divided by step width, input value for FSD or step
width will not be altered.
Dead load plus span is > 3.0 mV/V.
Span is < 0.3 mV/V for 3000 d (OIML or NTEP).
Span is < 0.03 mV/V for 3000 d, (W&M = none).
Span plus dead load is > 3.0 mV/V.