Hardware Configuration PR 5610 Installation Manual
26 Sartorius
3.3.4 RS 232 Built-in interface
The instrument is equipped with a standard RS 232 interface, called Builtin. If a PC with terminal program or a
terminal is connected, the interface can be used for calibration and configuration of the instrument. The
interface can be used freely and can be configured e.g. for data transmission to a supervisory system, a remote
display, a printer or an external Sartorius scale (xBPI).
Technical data
Connecting method: 1-row 5-pin connector Number of channels: 1
Type: RS 232 full duplex Speed: 300 to 115K2 Bit/sec
Data bits: 7 / 8 Bit Parity: None, odd, even
Number of signals: 2 output signals (TXD, RTS)
2 input signals (RXD, CTS)
Output signal level: logic 1 (high) - 5 ... - 15 V
logic 0 (low) + 5 ... + 15 V
Input signal level: logic 1 (high) - 3 ... - 15 V
logic 0 (low) + 3 ... + 15 V
Potential isolation: no
Cable type: twisted pairs, screened
(e.g. LifYCY 3x2x0.20),
1 pair of wires for GND
Cable length: max. 15 m
3.3.4.1 PC or terminal connection
The terminal
block for the
Builtin interface
is located on
board 1
If [Operator device at] is selected at [Builtin RS232], no further setup on the instrument is necessary, parameters
for the PC or terminal are: 9,600 bits/sec, 8 data bits, no parity, 1 stop bit and XON/XOFF protocol.
CX
2U-
"builtin"
RS232 Interface
CX
2U+
U / U
X602
CTS
GND
Tx
Rx
RTS
4
5
2
3
1
PR1713/..