PR 5610 Installation Manual Hardware Configuration
Sartorius 27
3.3.4.2 Remote display / remote terminal connection
The builtin RS 232 interface can be used to connect the remote display PR 1627 or the remote terminal PR 1628.
In [Serial port setup]-[Builtin RS232]-[Protocol] select [RemoteDsp], it has got some fixed parameters:
Bits-7, Parity-Even, Stopbits-1.
The [Baudrate] has to be set to [4800] to correspond with the default values of PR 1627 or PR 1628.
3.3.4.3 Ticket printer connection
The Ticket printer TM 295II/PH02 has got a special EPROM to run the [W&M Print] protocol.
The following setup has to be done: [Serial port setup]-[Builtin RS232]:[Protocol]-[W&M Print], [Baudrate]-
[4800], [Bits]-[7], [Parity]-[Even], [Stopbits]-[1], [Devtype]-[Raw] and [Echo]-[Disabled].
The printer has to be configured by the 10 DIP switches.