Hardware Configuration PR 5610 Installation Manual
54 Sartorius
3.4.11 PR 1721/11 Profibus interface
The Profibus interface PR 1721/11 is an option to be installed in slot 4 of PR 5610, see chapter 3.2.2.
Communication protocols and syntax comply with the relevant bus standard.
Connecting method: 50 pole connector, slot 4
Terminal block with 6 screw
terminals
Transmission rate:
Profibus-DP max. 12 Mbits/sec.
Protocols:
Profibus-DP slave
Cable: Shielded twisted pair
Cable impedance: 150 Ohm
Dimension (LxWxH): 70 x 86 x 25 mm Potential isolation: Yes
Weight: 65 grs.
Please ensure, that the two switches are set to 0 position
Connection diagram for a master with three slaves in Profibus
For further information please refer to the manuals Fieldbus-Interface PR 1721/1x (hardware) and