PR 5610 Installation Manual Hardware Configuration
Sartorius 29
3.4 Options
3.4.1 PR 1713/04 Serial I/O
The plug-in card contains 1x RS 232 and 1x RS 485/ 422, whereby only the RS 485 is galvanically isolated. Max.
3 cards PR 1713/04 can be used. The parameters of the RS 232 and RS 485/ 422 are adjusted by software
configuration. The RS 485/ RS 422 interface must be additionally configured by DIP switches on the card before
installation.
Connecting method: 2-row 25-pin connector
Terminal block A/B with 2x8
screw terminals
Number of channels: 1 RS 232, 1 RS 422/485
Type: full duplex
Speed: 300 to 19,200 Bit/sec
Signals RS 232: Output: TXD, RTS, DTR
Input: RXD, CTS, DCD, RI
Signals RS 422/485: TxA, RxA, TxB, RxB
Potential isolation: RS 232 no, RS 422 / 485 yes
Cable length: RS 232 max. 15m
RS 422 / 485 max. 1000m
Dimension (LxWxH): 80 x 68 x 15 mm
Weight: 33 grs.
Cable type: twisted pairs, screened
(e.g. LifYCY 3x2x0.20), 1 pair of
wires for GND.
Protocols are configurable by software set-up.
DIP switch S101 Settings for RS 422/ 485
1:
2:
3:
4:
5:
Tristate enable:
Rx:
pull-up resistor:
Bus terminator:
pull-down resistor
OFF - RS 422
OFF - continuously enabled
OFF - not conn.
OFF - not conn.
OFF - not conn.
ON - RS 485
ON - disabled while transm.
ON - (RxB 1k54 +V)
ON - (RxA 205E RxB)
ON - (RxA 1k54 -V)
The table indicates, which switches have to be closed (ON) for a certain mode.
PR 1713/04 Two-wire system Four-wire system
S101 point-to-point bus point-to-point bus
master RS 485
1, 2, 3, 4, 5 = on
RS 485
1, 2, 3, 4, 5 = on
RS 422
4 = on
RS 422
3, 4, 5 = on
single slave RS 485
1, 2 = on
- RS 422
4 = on
-
other slaves - RS 485
1, 2 = on
- RS 485
1 (default) = on
last slave - RS 485
1, 2, 3, 4, 5 = on
- RS 485
1, 3, 4, 5 = on