PR 5610 Installation Manual Hardware Configuration
Sartorius 43
3.4.6 PR 1713/12 4 Opto inputs / 4 Opto outputs
Max. 3 cards PR 1713/12 or PR 1713/15 can be used. The digital interfaces are passive
(external supply is
necessary) and have no common potential.
Connecting method: 2-row 25-pin connector
Terminal block A/B with 2x8
screw terminals
Number of inputs / outputs: Inputs: 4 , outputs: 4
Input signal:
Low: 0 ... 5V or open
High: 10 ... 31 V
External supply required
Input current: < 7 mA at 24 V
< 3 mA at 12 V
Output: Max. switching volt.: 32 V
Max. switching current : 25 mA
Voltage drop @ 25 mA: 3 V
External supply required
Dimension (LxWxH): 80 x 68 x 15 mm Potential isolation: Yes, optocoupled
Weight: 33 grs. Cable length (screened): Max. 50 m
• 4x outputs, dig. (On - Off) OUT for process control, galvanically isolated
with passive optocoupler outputs (open collector, max. 25 mA)
• 4x inputs, dig. (On - Off) IN for process control.
The inputs are galvanically isolated via optocoupler. The input signal of the digital interface is a logical "0" with
open input lines.
Accessories/ options: external supply required
Pin allocation of option connectors in the unit (installation see chapter 3.2.1)
Terminal block B input B1 B2 B3 B4 B5 B6 B7 B8
upper row 4 channels + CH1 - CH1 + CH2 - CH2 + CH3 - CH3 + CH4 - CH4
Terminal block A output A1 A2 A3 A4 A5 A6 A7 A8
lower row 4 channels + CH1 - CH1 + CH2 - CH2 + CH3 - CH3 + CH4 - CH4