PR 5610 Installation Manual Commissioning
Sartorius 75
Acknowledgement after pressing .
Continue with 4.5.10 Set full scale deflection (span).
4.5.10 Set full scale deflection (span)
The choices [Load], [mV/V] and [Data] for the span entry are
displayed.
[Load]
Select [Load], if the scale is to be loaded with a known weight.
Put the known weight on the scale, enter the weight value with
numeric keys, change the weight unit with
,
and confirm with
.
Acknowledgement after pressing .
The menu for calculating the test value is displayed. Continue
with 4.5.11 Calculate test value.
[mV/V]
Select [mV/V], if the span has been calculated or is known from a
previous calibration.
Enter span value in mV/V and confirm with .
Acknowledgement after pressing .
The menu for calculating the test value is displayed. Continue
with 4.5.11 Calculate test value.
[Data]
Select [Data], if the load cell data are known.
Select [Setup], [Enter or [Calc]. The menus are explained one after
the other.
[Setup]
After selection of [Setup] enter the number of load cells between
1 and 10 can with numeric keys or select with
or ; confirm
with
.
Enter the nominal capacity per load cell; confirm with
.