Commissioning with Terminal PR 5610 Installation Manual
138 Sartorius
5.6.11 GWT Scale
The weight is read directly from the PR 1720 or PR 5610. By means of a customized IEC 61131 program, the
weight value (read out with GET_WEIGHT) can be made available on the alphanumeric display.
The weighing point is indicated in the mask header:
Configuration of WP B
PR1713 Rel. 03.12 20.05.2004 09:05
Configuration of WP B
Type :GWT Scale
W & M : none
Overload : 9 d
Minwgt : 0 d
Update time : 300 ms
Fieldbus slave : 1
Serial number : 0
W
&
M
: select
none, OIML
or
NTEP
, see chapter 4.4.2
Enter maximum permissable
Overload
Permitted range: 0 … 9999999d
Enter weight limit for print-out
Minwgt
Permitted range: 0 … 9999999d
Enter
Update time
Permitted range: 50, 100...2000 ms
Error message with faulty cycle time entry
: invalid,
values greater 2000 will be set to 2000
Enter
Fieldbus slave
address
Permitted range: 1 … 32 (according to the position at the remote bus)
Enter board number
Serial number
If the board number (for PR 1720 rel. 2.50 is required) is not set to
0
, it will be checked. If the
number does not match, an error 15 is generated.
Finish the configuration with
E Exit configuration
If parameters were changed, the following prompt is displayed:
Save WP data (Y/N)