Commissioning PR 5610 Installation Manual
100 Sartorius
4.8.12 Enable download
This menu appears only if [Setup]-[Software Parameter]-[Software download] is set to [protected by setup].
Select access with .
The download is enabled, menu can be left with .
If the [Software download] is set to [protected by setup] and [Download enabled] is not active, an error
message appears on the PR 1750 (for Rel. 2.10: Download disabled, for Rel.< 2.10: only error no.) and the
download is blocked.
4.8.13 Reboot
Select access with .
Select [Cold] for cold start, [Warm] for warm start or [Bios] for
starting the Bios (booting).
After [Bios] select [Cold] for cold start, [Erase] for setting back
to default parameter or [Flash] for loading new Bios and/or
firmware and/or application software.
With the test function can be called. With test the functions
[ROM-Checksum], [Fast Ramtest], [Deep RAMtest], [SIL-NR-
Check], [EAROM-Check], [EAROM-Erase], [KBD-Check 32],
[KBD-Check 64] and [DSP-Check] (display) can be executed.
4.9 Config program for application parameters
The menu [Setup]-[Config] depends on the application package. For further information please refer to the
respective manual e.g. IBC Controller or FlowController .