PR 5610 Installation Manual Commissioning with Terminal
Sartorius 139
5.6.12 Sartorius scale (xBPI)
The weight is read from a Sartorius scale via the xBPI protocol.
The weighing point is indicated in the mask header:
Configuration of WP B
PR1713 Rel. 03.12 20.05.2004 09:05
Configuration of WP B
Type :xBPIScale
W & M : none
Overload : 9 d
Minwgt : 0 d
Update time : 300 ms
Serial number : 0
W
&
M
: select
none, OIML
or
NTEP
, see chapter 4.4.2
Enter maximum permissable
Overload
Permitted range: 0 … 9999999d
Enter weight limit for print-out
Minwgt
Permitted range: 0 … 9999999d
Enter
Update time
Permitted range: 50, 100...2000 ms
Error message with faulty cycle time entry
: invalid,
values greater 2000 will be set to 2000
Enter
Serial number
If serial number is not set to
0
, the scale . will be checked. If the number does not match, an error 15
is generated
Finish the configuration with
E Exit configuration
If parameters were changed, the following prompt is displayed:
Save WP data (Y/N)