PR 5610 Installation Manual Commissioning
Sartorius 69
4.4 Weighing point assignment
It has to be differenciated between physical weighing points (e.g. internal ADC, xBPI scale) and logical weighing
points (WP A, WP B and WP C). In menu [Setup]-[Weighingpoints]-[WP X]-[Assign] the physical weighing point
can be assigned to a logical weighing point according to the following table:
physical weighing point logical weighing point
A B C
[none]
<> <>
[internal]
<>
[liquid] / [user]
< > = default
[GWT-Scale]
[xBPIScale]
[A+B]
Note: A physical weighing point can be assigned only once to a logical weighing point.
Example: If [internal] is assigned to WP A, it cannot be assigned to WP B anymore.
[liquid] and [user] are treated as one physical weighing point, so either [liquid] or [user] can be assigned.
The weight display shows WP A as default, with
the weight display can be switched to WP B (toggle). If WP
C is set to [A+B], switches from WP A to WP B to WP C to WP A and so on.
Example for setting an xBPI scale as WP A and the internal ADC as WP B:
Select [Weighingpoints] with or ; confirm with .
Select displayed weighing point (WP A) with [Assign].
Select other function with or between [liquid],
[internal], [xBPIScale] or [GWT-Scale]; confirm with
.
Select other function for WP B with or between [liquid],
[internal], [xBPIScale] or [GWT-Scale]; confirm with
.
Select with or [none] or [A+B], confirm with .
Save with [YES], keep previous assignment with [NO].
Now [Setup] and [Config] for the xBPI-Scale can be carried out.
Select with or [WP B], than select with [Assign].
Select with or [Internal].
Confirm with .
Save with [YES], keep previous assignment with [NO].