Commissioning PR 5610 Installation Manual
68 Sartorius
Prepare the PC: FlashIt!32 must be started now on the connected PC. Select COM-Port, select firmware file
from the list in the window, press Flash button, wait for message completed.
If the message No communication No prompt from device appears on the screen, the
reason could be: wrong COM-Port selection or cable between PC and instrument are not or
wrong connected.
During flashing various status messages are displayed:
Subsequently, the boot menu with the following status line is displayed:
Initialize PR 5610 with [Cold].
Flash erases the databanks of the instrument, therefore Erase is not necessary before Flash, to erase internal
databanks.
If [Erase] is started, the safety question has to be answered before erase is executed, after that Reboot.