PR 5610 Installation Manual Commissioning
Sartorius 67
4.3.2.1 Weighing point initialization
After [Erase], [Cold], [Warm] or power-on the weighing points will be initialized.
This ensures that the physical weighing points are delivering valid weight values. If weighing points do not
deliver valid data, an error message is generated.
After [Erase] only the WP A will be initialized (see chapter 4.4 for WP assignment)
Example for power-on (WP A and WP B are assigned):
Firmware checksum is tested.
Warmstart is carried out.
Weighing point initialization, '?' appears during initialization,
'Ε' if WP is ready.
Both WPs are ready.
Error message if WPs are
not ready after 5 sec.
The weighing points which are not ready show the ERROR 9 message on the weight display (no communication).
After the weighing point initialization the instrument returns to the main menu.
4.3.2.2 Loading new firmware
The instrument includes firmware and application packages (e.g. Batch-X5). A firmware without application
package may be used for projects. You need the software FlashIt!32 Rel 2.20, if you want to replace the
application-PLC with your own PLC-project. Firmware and PLC program can be taken from the Complementary
SW-Tool box CD or from our internet site. Update the BIOS only if requested by the instrument.
Do not interrupt the flashing, because the loader program for FlashIt!32 is included in the BIOS,
otherwise the instrument needs to be repaired.
Settings of the interfaces: Connect the selected serial port with the PC. Port parameters are set
automatically during the flash procedure.
Prepare the PR 5610: Open menu [Setup]-[Reboot]-[Bios], wait for message Doing Bios..., Bios ready. Then
select [Flash], select the serial port:
Select the serial port, which will be used for downloading the
program.
If the serial port "Built-in" is occupied,
the interface can be selected with
or
(only available, if a serial I/O card PR 1713/04 is inserted).