Commissioning PR 5610 Installation Manual
66 Sartorius
[Cold] All programs and data loaded in RAM are erased: Depending on the application package these
could be recipes, component data etc. Data in EAROM remain unchanged.
Messages:
The firmware checksum is tested.
A cold start is carried out.
[Warm] All programs and data in RAM and EAROM remain unchanged.
Messages:
The firmware checksum is tested.
A warm start is carried out
[Erase] Calibration data are set to default. Licences are erased, software configuration, parameters for
Fieldbus, Network etc. are set to default.
Messages:
Erase is carried out
The firmware checksum is tested
A cold start is carried out.
[Test] Activate Test menu
Warning
The Deep RAMtest and the EAROM-Erase are destructive
(data are lost)
ROM-Test: BIOS passed, Firmware passed
Fast RAM-Test (size): RAM=040000+000000
Deep RAM-Test (destructive): Bank1= 262144 bytes, Bank2 = 0 bytes
SIL-NR-Check: Boardnr: 20060984
EAROM-Check: EAROM passed
EAROM-Erase (destructive): EAROM passed
KBD-Check: FFFDFFFF 0011 4031 (example: 1-key), end with +
DSP-Check: illuminate all segments, show characters with or ,
check weight display and status indicators with
or
[Flash] Loading of new firmware (refer to chapter 4.3.2.2).
Various messages are displayed in the boot menu:
Restart from the operating level was carried out
Loading of a new software was completed
Loading of a new software was cancelled
A fatal error occurred before restart
The checksum of the firmware is faulty
BIOS version not matching *
(BIOS too old after new firmware was flashed)
*) The BIOS must not be flashed without the message:
. Do not interrupt the
flashing, because the loader program for FlashIt!32 is included in the BIOS, otherwise the
instrument needs to be repaired.