PR 5610 Installation Manual Commissioning with Terminal
Sartorius 145
5.7.3.4 Test BCD-Output
The screen for testing PR 1713/08 is displayed, if the card is detected in the system. The actual input- and
output values are displayed binary and for the output also in hex- and decimal notation. Single output bits may
be changed by 'Set Bit' or 'Reset Bit'. You may enter output data in hex- or decimal format. By means of
'Control: X1Y0' the output is controlled by the input Enab.
5.7.3.5 Test DIOS Card
The screen for testing PR 1713/13 is displayed, if the card is detected in the system. The fields 'Ready' and 'Run'
must be active. Changes of the inputs at the remote bus are displayed. By means of 'Modify bytes' outputs at
the remote bus can be changed. The fields "BSA, CTR, RB, LB, MOD" are not in use, they are intended for future
extensions.
PR1713 Rel. 03.12 20.05.2004 09:05
Test PR1713/13 Slot-Number 3
Diag Bits=A0: Ready
Fail Run BSA CTR RB LB MOD
%IB 3.x =000100A001000000 IBS Status
%IB 3.1.x=0000 %QB 3.1.x=0000
E Exit Return Modify bits
5.7.3.6 Test Fieldbuscard
The screen for testing PR 1721 is displayed, if the card is detected in the system. Depending of the type of the
card a fitting headline is generated, additionally the input- and outputstring is displayed, e.g. for PR 1721/01:
PR1713 Rel. 03.12 20.05.2004 09:05
Test PR1721
PR1721 Profibus-DP Slave
Indata 0000000000000000
Outdata 0000000000000000
E Exit
5.7.4 Printing all configuration data
Calling up is done by pressing key [P]:
• Print out all configuration data:
P Print all configuration data
Printing out is done via the serial interface configured for the printer.