Loading...
Commissioning PR 5610 Installation Manual
74 Sartorius
4.5.7 Set full scale
Enter numeric values for FSD with or without decimal point.
Change weight unit with , confirm with .
Acknowledgement of full scale.
4.5.7.1 Resolution magnifier
The weight display can be switched to 10-fold during calibration with (toggle function).
4.5.8 Set step width
Select entry mode for step width with . The current value is
indicated.
Select the step width 1, 2, 5, 10, 20 or 50 with or and
confirm with
.
Acknowledgement of step width. Continue with 4.5.9 Set dead
load.
4.5.9 Set dead load
The choices [Load] and [mV/V] for the dead load entry are
displayed.
[Load]
Unload the scale and select the current weight as dead load with
[Load].
Calculation of dead load.
Continue with 4.5.10 Set full scale deflection (span).
[mV/V]
Select [mV/V], if dead load has been calculated or is known from
a previous calibration.
Enter dead load in mV/V and confirm with
.