Commissioning PR 5610 Installation Manual
80 Sartorius
4.6 Configuration for 'liquid', 'user', 'GWT-Scale', 'xBPI-Scale'
The physical weighing points can be assigned to the logical weighing points (WP A, WP B), see chapter 4.4.
After the assignment, the physical weighing points have to be configured. The configuration differs according
to the scale type. Therefore they explained one after the other. In all display clips the default value is displayed.
Select [Weighingpoints] with or ; confirm with .
With [Assign] the physical weighing point can be defined.
With or one of the options [liquid], [xBPI-Scale], [GWT-
Scale], [user] or [none] can be assigned and confirmed with
.
If the weighing point function has been changed, the change
can be stored with [Yes].
With [No] the weighing point function remains unchanged.
4.6.1 Weighing point function liquid counter
The weighing point function [liquid] needs a digital input for the pulses (counts). The fixed address is channel 6
(terminals B6-B7) of options PR 1713/17 plugged into position1 (slot 1).
With [Config] the configuration fof the liquid counter can be
accessed.
Select W
&M mode [none], [OIML], [NTEP] with or ; confirm
with
.
Enter numeric value for FSD with or without decimal point.
Change weight unit with ; confirm with .
Select the step width 1, 2, 5, 10, 20 or 50 with or and
confirm with
.
Enter the overload range > FSD with numeric keys from 0 to
9999999 d; confirm with .
Enter minimum weight for print-out with numeric keys from 0 to
9999999 d; confirm with .
Enter the increase in weight per x count(s) with numeric keys,
numeric range depends on FSD. Change the weight dimension
from [kg] to [t], [lb] or [g] with
; confirm with .
Enter the number of counts with numeric keys necessary to get
the defined weight increase, confirm with .
Enter time interval for weight data transfer from the external
weighing point with numeric keys; confirm with .