PR 5610 Installation Manual Commissioning with Terminal
Sartorius 119
5.3.1 Weighing point assignment
For general information refer to chapter 4.4.
Calling up mask
Assign weighingpoints
is done from the mask
Weighing point
by pressing key [A].
PR1713 Rel. 03.12 20.05.2004 09:05
Weighingpoint assignment
A :internal internal
B :GWT-Scale liquid
C :A+B user
(GWT-Scale)
xBPIScale
The physical weighing points (e.g. internal, xBPIScale) can be assigned to logical WPs (e.g. A, B):
Not defined
none
Liquid counter (internal)
liquid
Customer-specific weighing point
user
PR 1720 or PR 5610 / PR 1713 (via Interbus)
GWT Scale
Sartorius scale xBPI Scale
After entering [] the weighing point type for B can be selected using the two cursor keys [] and []. The
actually selected weighing point is highlighted by inverse display.
5.4 Weighing point parameters: storing, saving, protecting
This chapter describes how to save and protect parameters of the internal weighing point (internal ADC). The
individual weighing point parameters, entry, configuration and calibration are described in the two following
sub-sections.
5.4.1 Access protection: CAL-switch
The calibration data and parameters are protected against unauthorized/ accidental access by the CAL-switch
(on board 1). Check of the switch position is done when calling up mask
Calib/Config:
CAL-switch in open position: access permitted
When calling up mask
Calib/Config
, the weighing point can be configured and calibrated. In
addition to the actual weighing point parameters, mask
Calibration
indicates the access functions
which are enabled now (see chapter 5.4.1.1).
CAL-switch in closed position: calibration data are protected, download or flash not possible
When switching on the instrument
the calibration data saved in the EAROM (only with valid calibration data) are loaded.
otherwise, factory setting and factory calibration are loaded.
When calling up mask
Calib/Config
, the weighing point parameters are protected. The weighing point
parameters are only displayed in mask
Calib/Config
.
For W&M weighing systems, the CAL-switch has to be sealed in compliance with the calibration standards.